Validation of the Accuracy of Automatic Measurement of Blood Volume in Culture Bottles for Blood Culture

Author:

Kim Kyoungbo1ORCID,Park Sunggyun1ORCID

Affiliation:

1. Departments of Laboratory Medicine, Keimyung University School of Medicine, Daegu 42601, Republic of Korea

Abstract

Several manufacturers have developed systems that automatically measure the amount of blood in culture bottles. We compared the volumes measured automatically by the Virtuo instrument (bioMérieux, France) (height-based volumes) and those calculated by weighing the bottles. In all, 150 pairs of aerobic and anaerobic blood culture bottles (BacT/ALERT FA/FN Plus, bioMérieux) were randomly selected over two periods to compare the height- and weight-based volumes and analyze the effect of foam. We also estimated the limit of detection (LOD) and the cut-off value for 5 mL equine blood. The mean height-based volume was approximately 0.67 mL greater than the weight-based volume, particularly in anaerobic culture bottles. Foam did not have a significant effect. The LOD for the automatic height-based volume of equine blood was 0.2–0.4 mL. The 5 mL cut-off was 4–4.2 mL. Therefore, when reporting or monitoring blood volume within culture bottles in the laboratory, these performance characteristics should be adequately considered.

Publisher

MDPI AG

Subject

Clinical Biochemistry

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3