Affiliation:
1. Departments of Laboratory Medicine, Keimyung University School of Medicine, Daegu 42601, Republic of Korea
Abstract
Several manufacturers have developed systems that automatically measure the amount of blood in culture bottles. We compared the volumes measured automatically by the Virtuo instrument (bioMérieux, France) (height-based volumes) and those calculated by weighing the bottles. In all, 150 pairs of aerobic and anaerobic blood culture bottles (BacT/ALERT FA/FN Plus, bioMérieux) were randomly selected over two periods to compare the height- and weight-based volumes and analyze the effect of foam. We also estimated the limit of detection (LOD) and the cut-off value for 5 mL equine blood. The mean height-based volume was approximately 0.67 mL greater than the weight-based volume, particularly in anaerobic culture bottles. Foam did not have a significant effect. The LOD for the automatic height-based volume of equine blood was 0.2–0.4 mL. The 5 mL cut-off was 4–4.2 mL. Therefore, when reporting or monitoring blood volume within culture bottles in the laboratory, these performance characteristics should be adequately considered.