Abstract
This paper presents a method for gas concentration determination based on the measurement of the refractive index dispersion of a gas near the gas resonance in the near-infrared region (NIR). The gas refractive index dispersion line shape is reconstructed from the variation in the spectral interference fringes’ periods, which are generated by a low-finesse Fabry-Perot interferometer during the DFB diode’s linear-over-time optical frequency sweep around the gas resonance frequency. The entire sensing system was modeled and then verified experimentally, for an example of a low concentration methane-air mixture. We demonstrate experimentally a refractive index dispersion measurement resolution of 2 × 10−9 refractive index units (RIU), which corresponds to a change in methane concentration in air of 0.04 vol% at the resonant frequency of 181.285 THz (1653.7 nm). The experimental and modeling results show an excellent agreement. The presented system utilizes a very simple optical design and has good potential for the realization of cost-efficient gas sensors that can be operated remotely through standard telecom optical fibers.
Funder
Javna Agencija za Raziskovalno Dejavnost RS
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry