Abstract
A novel apparatus was developed, to investigate the detachment of particle structures consisting of soot and ash from a single fibre or a fibre array in hot gas flow. Key features of the novel apparatus are operation at high temperatures while two different measurement techniques are applied simultaneously in the same measurement chamber to observe particle structure detachment from a loaded fibre array. A heated inlet can heat the air stream at the position of the fibre array up to 470 °C, allowing detachment investigations at temperatures relevant for the operation of, e.g., soot particle filters. The first measurement technique integrated in the setup is video recording of the fibre array, which gives qualitative information on the rearrangement or detachment of particulate matter on the fibre. Because it is often difficult to distinguish rearrangement and detachment from pure visual observations, a second measurement technique is applied. This technique is a laser-light-sheet optical particle counter, which can detect detached particle structures and determine their size. The measurable size range is 257 to 1523 µm for glass spheres. This paper presents and discusses the novel apparatus, its calibration and first detachment measurement results.
Funder
Deutsche Forschungsgemeinschaft
Subject
Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry
Cited by
4 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献