Multi-Magnet Cochlear Implant Technology and Magnetic Resonance Imaging: The Safety Issue

Author:

Canzi Pietro12,Carlotto Elena3,Zanoletti Elisabetta4,Frijns Johan H. M.5ORCID,Borsetto Daniele6,Caruso Antonio7,Chiapparini Luisa8,Ciorba Andrea9ORCID,Conte Giorgio10ORCID,Creber Nathan6ORCID,Criscuolo Stefania8,Di Lella Filippo11,Franchella Sebastiano4ORCID,Hensen Erik F.5ORCID,Lauda Lorenzo7,Malpede Stefano1,Mandalà Marco12,Rotteveel Liselotte J. C.5,Simoncelli Anna8,Stellato Anna Chiara1,Zanetti Diego1314ORCID,Benazzo Marco12

Affiliation:

1. Department of Clinical, Surgical, Diagnostic and Pediatric Sciences, University of Pavia, 27100 Pavia, Italy

2. Department of Otorhinolaryngology, University of Pavia, Fondazione IRCCS Policlinico San Matteo, 27100 Pavia, Italy

3. Department of Otolaryngology, S. Croce Hospital, 12100 Cuneo, Italy

4. Section of Otorhinolaryngology-Head and Neck Surgery, Department of Neurosciences, University of Padova, 35127 Padova, Italy

5. Department of Otorhinolaryngology, Leiden University Medical Center, 2300 RC Leiden, The Netherlands

6. Department of ENT, Addenbrooke’s Hospital, Cambridge University Hospitals NHS Foundation Trust, Cambridge CB2 0QQ, UK

7. Department of Otology and Skull Base Surgery, Otologic Group, 29121 Piacenza, Italy

8. Department of Diagnostic Radiology and Interventional Radiology and Neuroradiology, University of Pavia, Fondazione IRCCS Policlinico San Matteo, 27100 Pavia, Italy

9. ENT and Audiology Department, University Hospital of Ferrara, 44122 Ferrara, Italy

10. Neuroradiology Department Fondazione IRCCS Ca’ Granda Ospedale Maggiore Policlinico Milan, 20122 Milan, Italy

11. Otolaryngology and Otoneurosurgery Unit, University of Parma, 43126 Parma, Italy

12. Department of Otology and Skull Base Surgery, Azienda Ospedaliera Universitaria Senese, 53100 Siena, Italy

13. Audiology Unit, Department of Specialistic Surgical Sciences Fondazione IRCCS Ca’ Granda, Ospedale Maggiore Policlinico Milan, 20122 Milan, Italy

14. Department of Clinical Sciences and Community Health, University of Milan, 20122 Milan, Italy

Abstract

Despite the spread of novel-generation cochlear-implant (CI) magnetic systems, access to magnetic resonance imaging (MRI) for CI recipients is still limited due to safety concerns. The aim of this study is to assess and record the experiences of Hires Ultra 3D (Advanced Bionics) recipients who underwent an MRI examination. A multicentric European survey about this topic was conducted focusing on safety issues, and the results were compared with the current literature. We collected a total of 65 MRI scans performed in 9 otologic referral centers for a total of 47 Hires Ultra 3D recipients, including, for the first time, 2 children and 3 teenagers. Preventive measures were represented by scanning time and sedation for children. Head wrapping was used in eight cases, and six of the eight cases received local anesthesia, even if both measures were not needed. Only three patients complained of pain (3/65 examinations, 4.6%) due to the tight head bandage, and one of the three cases required MRI scan interruption. No other adverse events were reported. We believe that these results should encourage MRI execution in accordance with manufacturer recommendations for Ultra 3D recipients.

Publisher

MDPI AG

Reference32 articles.

1. Magnetic Resonance Imaging in Patients with Cochlear Implants without Magnet Removal: A Radiology-Administered Protocol to Enhance Operational Efficiency and Improve Workflow;Johnson;Otol. Neurotol.,2023

2. Should patients with brain implants undergo MRI?;Erhardt;J. Neural. Eng.,2018

3. A Review of Reported Adverse Events in MRI-Safe and MRI-Conditional Cochlear Implants;Bestourous;Otol. Neurotol.,2022

4. Technavio (2024, February 03). Cochlear Implants Market by End-user, Product, and Geography—Forecast and Analysis 2023–2027. Available online: https://www.technavio.com/report/cochlear-implants-market-industry-analysis.

5. A review of the safety of MRI in cochlear implant patients with retained magnets;Srinivasan;Clin. Radiol.,2019