Can Steam Sterilization Affect the Accuracy of Point-of-Care 3D Printed Polyetheretherketone (PEEK) Customized Cranial Implants? An Investigative Analysis

Author:

Sharma Neha12ORCID,Zubizarreta-Oteiza Jokin12ORCID,Tourbier Céline12ORCID,Thieringer Florian M.12ORCID

Affiliation:

1. Clinic of Oral and Cranio-Maxillofacial Surgery, University Hospital Basel, 4031 Basel, Switzerland

2. Medical Additive Manufacturing Research Group (Swiss MAM), Department of Biomedical Engineering, University of Basel, Hegenheimermattweg 167C, 4123 Allschwil, Switzerland

Abstract

Polyetheretherketone (PEEK) has become the biomaterial of choice for repairing craniofacial defects over time. Prospects for the point-of-care (POC) fabrication of PEEK customized implants have surfaced thanks to the developments in three-dimensional (3D) printing systems. Consequently, it has become essential to investigate the characteristics of these in-house fabricated implants so that they meet the necessary standards and eventually provide the intended clinical benefits. This study aimed to investigate the effects of the steam sterilization method on the dimensional accuracy of POC 3D-printed PEEK customized cranial implants. The objective was to assess the influence of standard sterilization procedures on material extrusion-based 3D-printed PEEK customized implants with non-destructive material testing. Fifteen PEEK customized cranial implants were fabricated using an in-house material extrusion-based 3D printer. After fabrication, the cranial implants were digitalized with a professional-grade optical scanner before and after sterilization. The dimensional changes for the 3D-printed PEEK cranial implants were analyzed using medically certified 3D image-based engineering software. The material extrusion 3D-printed PEEK customized cranial implants displayed no statistically significant dimensional difference with steam sterilization (p > 0.05). Evaluation of the cranial implants’ accuracy revealed that the dimensions were within the clinically acceptable accuracy level with deviations under 1.00 mm. Steam sterilization does not significantly alter the dimensional accuracy of the in-house 3D-printed PEEK customized cranial implants.

Funder

Werner Siemens Foundation

Publisher

MDPI AG

Subject

General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3