Method Comparison of Erythrocyte Sedimentation Rate Automated Systems, the VES-MATIC 5 (DIESSE) and Test 1 (ALIFAX), with the Reference Method in Routine Practice

Author:

Cennamo Michele12,Giuliano Loredana2,Arrigoni Gloria2,Fardone Valentina2,Russo Roberta2,De Tomasi Luca Maria2,Bertani Fabio2ORCID,Cammarota Gaetano3,Brunetti Giovanni4,Del Vecchio Lucia5,Partenope Michelarcangelo2

Affiliation:

1. Department of Translational Medical Sciences, University of Naples “Federico II”, 80126 Naples, Italy

2. Clinical Pathology and Microbiology Unit, Laboratory Analysis, ASST Lariana, Hospital Sant’Anna, 22100 Como, Italy

3. Department of Diagnostic Innovation, Fondazione IRCCS Istituto Nazionale dei Tumori, 20133 Milan, Italy

4. Department of Diagnostics and Public Health, Section of Clinical Biochemistry, University of Verona, 37134 Verona, Italy

5. Department of Nephrology and Dialysis, ASST Lariana, Hospital Sant’Anna, 22100 Como, Italy

Abstract

Background: The erythrocyte sedimentation rate (ESR) is a routine and aspecific test that is still widely used. The reference-manual method for ESR determination is the Westergren method. The VES-MATIC 5 is a novel, fully automated, and closed system based on a modified Westergren method. This study conceived the aim of comparing two ESR analytical analysers, Test 1 and the VES-MATIC 5, with the reference method in routine practice. Methods: This study included 264 randomly analysed samples. A comparison between the two methods and Westergren was performed, and they were evaluated for inter-run and intra-run precision. In addition, we investigated possible interferences and different sensitivities to conventional analytes. Results: The comparison of methods by Passing–Bablok analysis provided a good agreement for both systems, with a better correlation for VES-MATIC 5 (p = 0.96) than Test 1 (p = 0.93), and sensitivity studies did not show any significant influence. Conclusions: The VES-MATIC 5 analyser demonstrated excellent comparability with the reference method, and it had better performance than Test 1. It can be employed in routine practice, bringing advantages such as a reduction in the probability of human error compared to the manual method, as well as an increase in operator safety and environmental protection.

Publisher

MDPI AG

Reference59 articles.

1. Tishkowski, K., and Gupta, V. (2023). StatPearls, StatPearls Publishing LLC.

2. Erythrocyte sedimentation in columns and the significance of ESR;Woodland;Biorheology,1996

3. The mechanism of erythrocyte sedimentation. Part 2: The global collapse of settling erythrocyte network;Pribush;Colloids Surf. B Biointerfaces,2010

4. The mechanism of erythrocyte sedimentation. Part 1: Channeling in sedimenting blood;Pribush;Colloids Surf. B Biointerfaces,2010

5. The erythrocyte sedimentation rate: Old and new clinical applications;Saadeh;South. Med. J.,1998

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献