Membrane artificial lungs (oxygenators) are used in cardiopulmonary surgery as well as, in some cases, in severe lung disease to support the natural lung by means of ECMO (extracorporeal membrane oxygenation). The oxygen and carbon dioxide transfer rates of any oxygenator are usually assessed by considering several blood gas parameters, such as oxygen saturation, hemoglobin concentration, partial pressure of oxygen and carbon dioxide, bicarbonate concentration, and pH. Here, we report a set of semi-empirical equations that calculate such parameters directly from their partial pressures and assess the acid–base balance during ECMO. The implementation of this equation set permits the evaluation of any oxygenator, existing or prototypes in development, as well as the development of clinical decision-making tools for predicting the blood gas state and acid–base balance during surgical interventions and ECMO. The predicted results are then compared with experimental data obtained from in vitro gas exchange investigations with a commercial oxygenator using fresh porcine blood. The high correlation, R2>0.95, between the predicted and the experimental data suggests a possibility of using such empirical equations in the simulation of gas transfer in a cardiopulmonary system with an oxygenator for any venous inlet blood gas data and also for estimating the acid–base balance during such therapy.
Funder
European Regional Development Funds
Subject
General Mathematics,Engineering (miscellaneous),Computer Science (miscellaneous)
Reference12 articles.
1. Nosé, Y. (1973). Manual on Artificial Organs, Mosby. The oxygenator.
2. Development of the oxygenator: Past, present, and future;Iwahashi;J. Artif. Organs,2004
3. Extracorporeal CO2-removal: Pulmonary and extracorporeal equilibria in dogs and sheep;Mottaghy;Eur. J. Appl. Physiol. Occup. Physiol.,1985
4. Mottaghy, K., Steyrer, K., Freund, E., and Dringenberg, I. (1978). Eine standardisierte Methode zur Bestimmung der Gasaustauschkapazität der Oxygenatoren am Beispiel eines Bubble -und Liquid- Oxygenators. Biomed. Tech., 23.
5. (2016). Cardiovascular Implants, and Artificial Organs—Blood-Gas Exchangers (Oxygenators). Standard No. ISO 7199:2016.
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
1. Mathematical modeling of oxygen and carbon dioxide exchange in hollow fiber oxygenators;Up-to-Date Approach to Blood Gas Analysis – Clues to Diagnosis and Treatment [Working Title];2024-09-02