Continuous Non-Invasive Blood Pressure Measurement Using 60 GHz-Radar—A Feasibility Study

Author:

Vysotskaya Nastassia12ORCID,Will Christoph1ORCID,Servadei Lorenzo3ORCID,Maul Noah2ORCID,Mandl Christian1,Nau Merlin2ORCID,Harnisch Jens1,Maier Andreas2ORCID

Affiliation:

1. Infineon Technologies AG, Am Campeon 1-15, 85579 Neubiberg, Germany

2. Department for Computer Science 5 (Pattern Recognition), Friedrich-Alexander-University Erlangen-Nuremberg (FAU), Martensstrasse 3, 91058 Erlangen, Germany

3. Department of Electrical and Computer Engineering, Technical University of Munich, Arcisstrasse 21, 80333 Munich, Germany

Abstract

Blood pressure monitoring is of paramount importance in the assessment of a human’s cardiovascular health. The state-of-the-art method remains the usage of an upper-arm cuff sphygmomanometer. However, this device suffers from severe limitations—it only provides a static blood pressure value pair, is incapable of capturing blood pressure variations over time, is inaccurate, and causes discomfort upon use. This work presents a radar-based approach that utilizes the movement of the skin due to artery pulsation to extract pressure waves. From those waves, a set of 21 features was collected and used—together with the calibration parameters of age, gender, height, and weight—as input for a neural network-based regression model. After collecting data from 55 subjects from radar and a blood pressure reference device, we trained 126 networks to analyze the developed approach’s predictive power. As a result, a very shallow network with just two hidden layers produced a systolic error of 9.2±8.3 mmHg (mean error ± standard deviation) and a diastolic error of 7.7±5.7 mmHg. While the trained model did not reach the requirements of the AAMI and BHS blood pressure measuring standards, optimizing network performance was not the goal of the proposed work. Still, the approach has displayed great potential in capturing blood pressure variation with the proposed features. The presented approach therefore shows great potential to be incorporated into wearable devices for continuous blood pressure monitoring for home use or screening applications, after improving this approach even further.

Funder

German Federal Ministry of Education and Research

Publisher

MDPI AG

Subject

Electrical and Electronic Engineering,Biochemistry,Instrumentation,Atomic and Molecular Physics, and Optics,Analytical Chemistry

Cited by 8 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Enhanced Discrete Wavelet Transform–Non-Local Means for Multimode Fiber Optic Vibration Signal;Photonics;2024-07-07

2. Radar-Based Human Pulse Sensing through a Resonator-Based Superstrate;2024 IEEE MTT-S International Microwave Biomedical Conference (IMBioC);2024-06-11

3. A Non-Invasive and Non-Contact Jugular Venous Pulse Measurement: A Feasibility Study;2024-06-07

4. UWB-enabled Sensing for Fast and Effortless Blood Pressure Monitoring;Proceedings of the ACM on Interactive, Mobile, Wearable and Ubiquitous Technologies;2024-05-13

5. Transforming Cardiovascular Health: a Transformer-Based Approach to Continuous, Non-Invasive Blood Pressure Estimation via Radar Sensing;ICASSP 2024 - 2024 IEEE International Conference on Acoustics, Speech and Signal Processing (ICASSP);2024-04-14