Author:
Khalil Sabry,El-Sharnouby Mohamed
Abstract
A highly Co (II) liquid ion-selective electrode depending on the reaction of cobalt ions with the reagent 2-(5-Bromo-2-pyridylazo)-5-[N-n-propyl-N-(3-sulfopropyl) amino] aniline is successfully fabricated. The characteristic slope (56.66 mV), the linear range response from 3.4 × 10−8 to 2.4 × 10−2 molar, the detection limit (2.7 × 10−8) molar, the selectivity coefficient toward some metal ions, the time of response (10 s), lifetime (seven months), the pH effect on the sensor potential and the basic analytical parameters were studied. The sensor was used to estimate the concentration of cobalt ions in food products and pharmaceutical formulations. The obtained results of the developed sensor were statistically analyzed and compared with those of other different reported electrodes.
Subject
Physical and Theoretical Chemistry,Analytical Chemistry
Reference27 articles.
1. Intolerability of cobalt salt as erythropoietic agent
2. Al(III)-selective electrode based on newly synthesized xanthone derivative as neutral ionophore
3. Cobalt use and regulation in horseracing;Kalayci;Talanta,2005
4. Construction of a PVC based 15-crown-5 electrochemical sensor for Ag(I) cation
5. Spectrophotometric determination of cobalt in biological and environmental samples using 2, 6-pyridinedicarboxaldehyde thiosemicarbazone;Ahmed;Int. J. Anal. Pharm. Bio. Sci.,2013
Cited by
9 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献