Abstract
The purpose of this study was to evaluate the change in the retentive forces of four different titanium-based implant attachment systems during the simulation of insert–removal cycles in an artificial oral environment. Five types of titanium-based dental implant attachment systems (Locator, Kerator, O-ring, EZ-Lock, and Magnetic) were studied (n = 10). The specimens underwent insert–removal cycles in artificial saliva, and the retentive force was measured following 0, 750, 1500, and 2250 cycles. Significant retention loss was observed in all attachment systems, except the magnetic attachments, upon completion of 2250 insertion and removal cycles, compared to the initial retentive force (p < 0.05). A comparison of the initial retentive forces revealed the highest value for Locator, followed by the Kerator, O-ring, EZ-Lock, and Magnetic attachments. Furthermore, Kerator demonstrated the highest retentive loss, followed by Locator, O-ring, EZ-Lock, and Magnetic attachments after 2250 cycles (p < 0.05). In addition, the Locator and Kerator systems revealed significant decrease in retentive forces at all measurement points (p < 0.05). The retention force according to the insert–removal cycles were significantly different according to the types of dental implant attachment systems.