Technical note: Energy dependence of the Gafchromic EBT4 film: Dose‐response curves for 70 kV, 6 MV, 6 MV FFF, 10 MV FFF, and 15 MV x‐ray beams

Author:

Chan Maria F.1,Park Jeonghoon1,Aydin Resat2,Lim Seng‐Boh1

Affiliation:

1. Department of Medical Physics Memorial Sloan Kettering Cancer Center New York New York USA

2. Ashland Inc. Bridgewater New Jersey USA

Abstract

AbstractBackgroundEBT4 was newly released for radiotherapy quality assurance to improve the signal‐to‐noise ratio in radiochromic film dosimetry. It is important to know its dose‐response characteristics before its use in the clinic.PurposeThis study aims to investigate and compare the dose‐response curves of the Gafchromic EBT4 film for megavoltage and kilovoltage x‐ray beams with different dose levels, scanning spatial resolutions, and sizes of region of interest (ROI).MethodsEBT4 film (Lot#07052201) calibration strips (3.5 × 20 cm2) were exposed to a 10×10 cm2 open field at doses of 0, 63, 125, 500, 750, 1000 cGy using 6 MV photon beam. EBT4 film strips from the same lot were then exposed to each x‐ray beam (6 MV, 6 MV FFF, 10 MV FFF, 15 MV, and 70 kV) at six dose values (50, 100, 300, 600, 800, 1000 cGy). A full sheet (25 × 20 cm2) of EBT4 film was irradiated at each energy with 300 cGy for profile comparison with the treatment planning calculation. At two different spatial resolutions of 72 and 300 dpi, each film piece was scanned three consecutive times in the center of an Epson 10000XL flatbed scanner in 48‐bit color. The scanned images were analyzed using FilmQA Pro. For each scanned image, an ROI of 2 × 2 cm2 at the field center was selected to obtain the average pixel value with its standard deviation in the ROI. An additional ROI of 1 cm diameter circle was also used to evaluate the impact of ROI shape and size, especially for FFF beams. The dose value, average dose‐response value, and associated uncertainty were determined for each energy and relative responses were analyzed. The Student's t‐test was performed to evaluate the statistical significance of the dose‐response values with different color channels, ROI shapes, and spatial resolutions.ResultsThe dose‐response curves for the five x‐ray energies were compared in three color channels. Weak energy dependence was found among the megavoltage beams. No significant differences (average ∼1.1%) were observed for all doses in this study among 6 MV, 6 MV FFF, 10 MV FFF, and 15 MV beams, regardless of spatial resolution and color channel. However, a statistically significant difference in dose‐response was observed up to 12% between 70 kV and 6 MV beams.ConclusionsThe dose‐response curves for Gafchromic EBT4 films were nearly independent of the energy of the photon beams among 6 MV, 6 MV FFF, 10 MV FFF, and 15 MV. For very low‐energy photons (e.g., 70 kV), a separate calibration from the same low‐energy x‐ray is necessary.

Publisher

Wiley

Subject

General Medicine

Cited by 4 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3