Assessing quality assurance of multi‐leaf collimator using the structural similarity index

Author:

Zhang Hong1,Zhang Baoshe2,Lasio Giovanni2,Chen Shifeng2,Nasehi Tehrani Joubin2

Affiliation:

1. Departments of Radiation Oncology Vanderbilt University Medical Center Nashville Tennessee USA

2. Departments of Radiation Oncology Medical School University of Maryland Baltimore Maryland USA

Abstract

AbstractPurposeThis study aims to evaluate the viability of utilizing the Structural Similarity Index (SSI*) as an innovative imaging metric for quality assurance (QA) of the multi‐leaf collimator (MLC). Additionally, we compared the results obtained through SSI* with those derived from a conventional Gamma index test for three types of Varian machines (Trilogy, Truebeam, and Edge) over a 12‐week period of MLC QA in our clinic.MethodTo assess sensitivity to MLC positioning errors, we designed a 1 cm slit on the reference MLC, subsequently shifted by 0.5–5 mm on the target MLC. For evaluating sensitivity to output error, we irradiated five 25 cm × 25 cm open fields on the portal image with varying Monitor Units (MUs) of 96–100. We compared SSI* and Gamma index tests using three linear accelerator (LINAC) machines: Varian Trilogy, Truebeam, and Edge, with MLC leaf widths of 1, 0.5, and 0.25 mm. Weekly QA included VMAT and static field modes, with Picket fence test images acquired. Mechanical uncertainties related to the LINAC head, electronic portal imaging device (EPID), and MLC during gantry rotation and leaf motion were monitored.ResultsThe Gamma index test started detecting the MLC shift at a threshold of 4 mm, whereas the SSI* metric showed sensitivity to shifts as small as 2 mm. Moreover, the Gamma index test identified dose changes at 95MUs, indicating a 5% dose difference based on the distance to agreement (DTA)/dose difference (DD) criteria of 1 mm/3%. In contrast, the SSI* metric alerted to dose differences starting from 97MUs, corresponding to a 3% dose difference. The Gamma index test passed all measurements conducted on each machine. However, the SSI* metric rejected all measurements from the Edge and Trilogy machines and two from the Truebeam.ConclusionsOur findings demonstrate that the SSI* exhibits greater sensitivity than the Gamma index test in detecting MLC positioning errors and dose changes between static and VMAT modes. The SSI* metric outperformed the Gamma index test regarding sensitivity across these parameters.

Publisher

Wiley

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. A novel EPID‐based MLC QA method with log files achieving submillimeter accuracy;Journal of Applied Clinical Medical Physics;2024-06-20

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3