Dosimetric effects of rotational errors for single isocenter multiple targets in HyperArc plans: A phantom and retrospective imaging analysis study

Author:

Golmakani Samane1,McGrath Andrew N.1,Williams Timothy J.1

Affiliation:

1. W.P. Holman Clinic Royal Hobart Hospital Hobart Tasmania Australia

Abstract

AbstractPurposeThis study uses a phantom to investigate the dosimetric impact of rotational setup errors for Single Isocenter Multiple Targets (SIMT) HyperArc plans. Additionally, it evaluates intra‐fractional rotational setup errors in patients treated with Encompass immobilization system.MethodsThe Varian HyperArc system (Varian Medical systems) was used to create plans targeting spherical PTVs with diameters of 5, 10, and 15 mm and with offsets of 1.3–5.3 cm from the isocenter. Dosimetric parameters, including mean and maximum dose, D99% and D95% were evaluated for various rotational setup errors ranging from 0.5° to 2° for the PTVs and certain CTVs created within PTVs. These rotational errors were applied in an order and direction that resulted in the maximum displacement of targets. The rotation was applied both uniformly around all three axes and individually around each axis. Furthermore, to link the findings to actual treatment scenarios, the intra‐fractional rotational setup errors were obtained for stereotactic cranial patients treated with the Encompass system using CBCT images acquired during treatments.ResultsThe maximum displacement of 2.7 mm was observed for targets located at 4.4 and 4.5 cm from the isocenter with rotational setup errors of 2°. The dose reduction for D99% values corresponding to this displacement were about 50%, 40%, and 30% for PTVs with diameters of 5, 10, and 15 mm, respectively. Both D99% and D95% showed a consistent trend of dose reduction across various rotational errors and PTV volumes. While the maximum dose remained consistent for different targets with various rotational errors, the mean dose decreased by approximately 25%, 12%, and 6% for PTVs with diameters of 5, 10, and 15 cm, respectively, with rotational errors of 2°. In addition, by analyzing CBCT images, the absolute mean rotational setup errors obtained during treatment with Encompass for pitch, roll, and yaw were 0.17° ± 0.13°, 0.11° ± 0.10°, and 0.12° ± 0.10° respectively. This data, combined with existing studies, suggest that a 0.5° rotational setup error is a safe choice to consider for calculating additional PTV margin to ensure adequate CTV coverage.Therefore, the assessment of maximum displacement and dosimetric parameters in this study, for a 0.5° rotational error, highlights the need for an additional 0.7 mm PTV margin for targets positioned at distances of 4.4 cm or greater from the isocenter.ConclusionsFor SIMT Plans, a 0.5° rotational setup error is recommended as a basis for calculating additional PTV margins to ensure adequate CTV coverage when using the Encompass system.

Publisher

Wiley

Subject

Radiology, Nuclear Medicine and imaging,Instrumentation,Radiation

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3