Abstract
AbstractBackgroundWhile most VVI pacemakers in bradycardic patients are set to a low limit of 60/min, the optimal lower limit rate for VVI pacemakers in atrial fibrillation has not been established. Although an increase in heart rate within the normal range in the setting of a VVI pacemaker might be expected to lead to an increase in cardiac output with the shortening of the diastolic time, the changes in cardiac output at different pacemaker settings have not been fully clarified.MethodsWe included 11 patients with bradycardic atrial fibrillation who had VVI pacemakers implanted. Stroke volume was measured using the electrical cardiometry method (AESCULONⓇ mini; Osypka Medical) without pacing and at ventricular pacings of 60, 70, 80, and 90/min.ResultsStroke volume decreased stepwise at ventricular pacing rates of 60, 70, 80, and 90/min (63.6 ± 11.2, 61.9 ± 10.6, 59.3 ± 12.2, and 57.5 ± 12.2 mL, p < .001), but cardiac output increased (3.81 ± 0.67, 4.33 ± 0.74, 4.74 ± 0.97, and 5.17 ± 1.09 L/min, p < .001). The rate of increase in cardiac output at a pacing rate of 70/min compared to 60/min correlated with left ventricular end‐systolic volume (r = 0.711, p = .014).ConclusionsCardiac output increased at a pacing rate of 70 compared to 60 in bradycardic atrial fibrillation patients, and the rate of increase in cardiac output was greater in those with larger left ventricular end‐systolic volume.