Affiliation:
1. UCLA David Geffen School of Medicine Department of Head and Neck Surgery Los Angeles California USA
Abstract
AbstractBackgroundTransoral robotic surgery (TORS) using the Single‐Port system (SPS) relies on electrocautery, limiting its applications in the upper aerodigestive tract. We evaluated the feasibility of a CO2 delivery system for the SPS.MethodsOtolaryngology residents performed a cutting exercise using a handheld CO2 laser and participated in a cadaveric oropharyngeal dissection using the SPS with monopolar cautery (SP + EC) and CO2 laser (SP + CO2). Residents completed the System Usability Scale (SUS) questionnaire to evaluate these techniques.ResultsThe same laser fiber was used for all combined dissections. The handheld CO2 laser, SP + EC, and SP + CO2 demonstrated similar SUS scores. On individual domain scores, SP + CO2 received less favorable ratings compared to the handheld CO2 laser for complexity, integration, and cumbersome experience (p < 0.05). On subgroup stratification, less TORS experience was associated with worse SUS scores.ConclusionSP‐guided CO2 laser delivery is a viable alternative to electrocautery in robotic surgery, and should be considered when performing TORS.
Subject
Computer Science Applications,Biophysics,Surgery
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献