Research on the correction method for radiotherapy verification plans based on displaced electronic portal imaging device

Author:

Guo Jian1ORCID,Zhou Leyuan2,Zeng Haibin2ORCID

Affiliation:

1. Department of Radiation Oncology The First Affiliated Hospital of Soochow University Suzhou China

2. Department of Radiation Oncology The Fourth Affiliated Hospital of Soochow University Suzhou China

Abstract

AbstractBackgroundIt has been observed that under the single isocenter conditions, the potential shifts of the electronic portal imaging devices (EPID) may be introduced when executing portal dosimetry (PD) plans for bilateral breast cancer, pleural mesothelioma, and lymphoma. These shifts are relative to the calibration positions of EPID and result in significant discrepancies in the plan verification results.PurposeTo explore methods including correction model and specific correction matrices to revise the data obtained from displaced EPID.MethodsTwo methods, the correction model and the specific correction matrices, were applied to correct the data. Five experiments were designed and conducted to build correction model and to validate the effectiveness of these two methods. Gamma passing rates were calculated and data profiles along X‐axis and Y‐axis were captured.ResultsThe gamma passing rates for the EPID‐displaced IMRT validation plans after applying correction model, along with the application of specific correction matrices to VMAT and IMRT validation plans, exhibit results that are comparable to the cases with non‐displaced EPID. Except for the VMAT plans applied correction model which showed larger discrepancies (0.041 ± 0.028, 0.049 ± 0.030), the other three exhibit minimal differences in discrepancy values. In all profiles, the corrected data from displaced EPID exhibit a high level of agreement with data obtained from non‐displaced EPID. Good consistency is observed in actual application of the correction model and the specific correction matrices between gamma passing rates of data corrected and those of non‐displaced data.ConclusionsThe proposed methods involving correction model and specific correction matrices can correct the data collected from the displaced EPID, and the gamma passing rates of the corrected data show results that are comparable to some extent with those of non‐displaced data. Particularly, the results corrected by specific correction matrices closely resemble the data from non‐displaced EPID.

Publisher

Wiley

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3