Technical note: Determination of CQ$C_{Q}$ for a miniature x‐ray source using a soft x‐ray ionization chamber calibrated in NIST reference beam qualities

Author:

Watson Peter G. F.1,Davis Stephen2,Culberson Wesley S.3

Affiliation:

1. Medical Physics Unit McGill University Montreal Quebec Canada

2. Department of Radiation Oncology Miami Cancer Institute Miami Florida USA

3. Department of Medical Physics School of Medicine and Public Health University of Wisconsin‐Madison Madison Wisconsin USA

Abstract

AbstractBackgroundThe formalism proposed by Watson et al. allows users of the INTRABEAM (Carl Zeiss Medical AG, Jena, Germany) electronic brachytherapy system to accurately determine the absorbed dose to water, in the absence of a primary dosimetry standard. However, all published values are for PTW 34013 ionization chambers calibrated in a TW30 reference beam, traceable to PTB (Germany). For North American users, it would be advantageous to have data for chambers calibrated in a kV reference beam maintained by the National Institute of Standards and Technology (NIST).PurposeIn this work, we determine for a PTW 34013 chamber calibrated in three NIST‐traceable reference beams: M30, L40, and L50.MethodsUsing available photon spectra data for M30, L40, and L50 reference beam qualities, Monte Carlo simulations using EGSnrc were performed to calculate the ratio of the absorbed dose to the PTW 34013 chamber air cavity to air‐kerma () for these beams. From this ratio, as a function of depth in water was determined. The effect of the use of a buildup foil was also investigated. An uncertainty analysis considering both the Type A and Type B uncertainties in the calculation of was performed.ResultsThe largest difference in was found between L50 and TW30, with a relative decrease of 1.4% (no buildup) to 1.6% (buildup). For M30 and L40, the differences were minimal compared with measurement uncertainties.ConclusionsWe report values for three NIST‐traceable kV reference beams. This study reinforces the feasibility of adapting the Watson et al. methodology using different kV reference beams, facilitating the use of INTRABEAM in North America and ensuring the continuity and accuracy of dosimetry standards in intraoperative radiation therapy.

Publisher

Wiley

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3