Simultaneous T2‐weighted real‐time MRI of two orthogonal slices

Author:

Hickey Samantha1ORCID,Reichert Andreas1ORCID,Ptacek Wolfgang2,Bielak Lars13ORCID,Reiss Simon1ORCID,Fischer Johannes1ORCID,Gunashekar Deepa Darshini1ORCID,Bortfeld Thomas4,Bock Michael1ORCID

Affiliation:

1. Division of Medical Physics, Department of Diagnostic and Interventional Radiology, University Medical Center Freiburg, Faculty of Medicine University of Freiburg Freiburg Germany

2. ACMIT Gmbh Wiener Neustadt Austria

3. German Cancer Consortium (DKTK) Partner Site Freiburg Freiburg Germany

4. Division of Radiation Biophysics, Department of Radiation Oncology Massachusetts General Hospital and Harvard Medical School Boston Massachusetts USA

Abstract

AbstractPurposeMR guidance is used during therapy to detect and compensate for lesion motion. T2‐weighted MRI often has a superior lesion contrast in comparison to T1‐weighted real‐time imaging. The purpose of this work was to design a fast T2‐weighted sequence capable of simultaneously acquiring two orthogonal slices, enabling real‐time tracking of lesions.MethodsTo generate a T2 contrast in two orthogonal slices simultaneously, a sequence (Ortho‐SFFP‐Echo) was designed that samples the T2‐weighted spin echo (S) signal in a TR‐interleaved acquisition of two slices. Slice selection and phase‐encoding directions are swapped between the slices, leading to a unique set of spin‐echo signal conditions. To minimize motion‐related signal dephasing, additional flow‐compensation strategies are implemented. In both the abdominal breathing phantom and in vivo experiments, a time series was acquired using Ortho‐SSFP‐Echo. The centroid of the target was tracked in postprocessing steps.ResultsIn the phantom, the lesion could be identified and delineated in the dynamic images. In the volunteer experiments, the kidney was visualized with a T2 contrast at a temporal resolution of 0.45 s under free‐breathing conditions. A respiratory belt demonstrated a strong correlation with the time course of the kidney centroid in the head–foot direction. A hypointense saturation band at the slice overlap did not inhibit lesion tracking in the semi‐automatic postprocessing steps.ConclusionThe Ortho‐SFFP‐Echo sequence delivers real‐time images with a T2‐weighted contrast in two orthogonal slices. The sequence allows for simultaneous acquisition, which could be beneficial for real‐time motion tracking in radiotherapy or interventional MRI.

Publisher

Wiley

Subject

Radiology, Nuclear Medicine and imaging

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献