Accuracy evaluation of Roche Accu‐Chek Performa blood glucose meters at low glucose concentrations: A nine‐year retrospective study

Author:

Zhao Zhipeng1,Li Runqing1ORCID,Zhao Xiuying1,Zhang Lina2,Wang Tengjiao1,Yang Song1,Han Ning1,Zhu Dong1

Affiliation:

1. Laboratory Medicine Department of Beijing Tsinghua Changgung Hospital School of Clinical Medicine Tsinghua University Beijing China

2. Operations Management Department of Beijing Tsinghua Changgung Hospital School of Clinical Medicine Tsinghua University Beijing China

Abstract

AbstractObjectiveTo evaluate the accuracy of Roche Accu‐Chek Performa glucose meters at a low glucose concentration of <5.55 mmol/L (100 mg/dL) over a 9‐year period.MethodsThe accuracy of the Roche Accu‐Chek Performa glucose meters at low glucose concentrations was evaluated using annual comparison data for 9 consecutive years from 2015 to 2023, according to the acceptability criteria specified in International Organization for Standardization (ISO) 15197:2013. Blood samples with low glucose concentrations of <5.55 mmol/L were prepared by incubation and glycolysis. The glucose concentration was detected using Roche Accu‐Chek Performa glucose meters and a biochemical analyzer in the central laboratory.ResultsA total of 2978 pairs of comparison results from 211 glucose meters at a low glucose concentration of <5.55 mmol/L were retrospectively analyzed from 2015 to 2023. The clinical use duration spanned from 1 to 9 years and 40.76% (86 out of 211 glucose meters) had been used for more than 2 years. The correlation coefficient r between glucose meter measurements and laboratory reference values was 0.98 (p < 0.001). The mean according to Roche Accu‐Chek Performa glucose meters was 0.05 mmol/L (0.9 mg/dL) higher than that of the biochemical analyzer (Z = −13.82, p < 0.0001). The results showed that 100.00% (211 out of 211) of the Roche Accu‐Chek Performa glucose meters met the acceptability criteria specified in ISO 15197:2013. At a low glucose concentration of <5.55 mmol/L, 99.90% (2975 out of 2978) of the comparative data pairs in the error distribution fell within the range of ±0.83 mmol/L (15 mg/dL). Parkes consensus error grid analysis showed that 100.00% (2978 out of 2978) of comparative data pairs fell within region A.ConclusionsThis study demonstrated that Roche Accu‐Chek Performa glucose meters successfully met the accuracy standards of ISO 15197:2013 for measuring blood glucose within the hypoglycemic range. Greater attention should be given to the performance of blood glucose monitoring systems in the low glycemic range, especially for patients with diabetes who are prone to hypoglycemia and require precise measurements.

Publisher

Wiley

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3