Affiliation:
1. University of Arizona College of Medicine Tucson Arizona USA
2. Mayo Clinic Arizona Phoenix Arizona USA
Abstract
AbstractPurposeTo determine whether a 24 MHz transducer significantly improves sensitivity, specificity and accuracy in evaluating flexor tendon repair integrity compared with an 18 MHz transducer.MethodsOne hundred and twelve cadaveric digits were randomised to an intact repair or simulated ‘failed’ repair, and to a two‐ or eight‐strand repair of a flexor digitorum profundus laceration. A blinded sonologist evaluated specimens in static mode using 18 and 24 MHz transducers. Gaps were remeasured after scanning, and final gap width recorded. McNemar's exact test calculated differences between sensitivity, specificity and accuracy, and chi‐squared test to compare sensitivity, specificity and accuracy between number of strands (2 vs. 8) and repair gap (≥4 mm).ResultsThe 24 MHz transducer had higher sensitivity (81 vs. 59%), lower specificity (67 vs. 70%) and higher overall accuracy (74 vs. 64%), than the 18 MHz transducer. The difference for sensitivity was significant (P = 0.011), but not differences for specificity and overall accuracy (P > 0.05). Pearson's correlation (r = 0.61) demonstrated a moderate‐to‐strong positive correlation between measured and true gap sizes. Increased number of suture strands (2 vs. 8) did not impair sensitivity, specificity nor accuracy.DiscussionUltrasound may tend to overestimate gap width, and a slight risk that some intact repairs, or those with small, clinically insignificant gaps may undergo surgical exploration that may not be indicated.ConclusionsA 24 MHz transducer is a more sensitive and accurate transducer for assessing flexor tendon repair integrity and measuring small gaps.
Subject
Radiology, Nuclear Medicine and imaging,Radiological and Ultrasound Technology
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
1. Multifaceted uses for musculoskeletal ultrasound;Australasian Journal of Ultrasound in Medicine;2023-11