A New Blood Pulsation Simulator Platform Incorporating Cardiovascular Physiology for Evaluating Radial Pulse Waveform

Author:

Yang Tae-Heon1ORCID,Kim Jaeuk U.2ORCID,Kim Young-Min2,Koo Jeong-Hoi3,Woo Sam-Yong4ORCID

Affiliation:

1. Department of Electronic Engineering, Korea National University of Transportation, Chungju-si 27469, Republic of Korea

2. Future Medicine Division, Korea Institute of Oriental Medicine, Daejeon 34054, Republic of Korea

3. Department of Mechanical and Manufacturing Engineering, Miami University, Oxford 45056, Ohio, USA

4. Center for Mechanical Metrology, Korea Research Institute of Standards and Science, Daejeon 34113, Republic of Korea

Abstract

To meet the need for “standard” testing system for wearable blood pressure sensors, this study intends to develop a new radial pulsation simulator that can generate age-dependent reference radial artery pressure waveforms reflecting the physiological characteristics of human cardiovascular system. To closely duplicate a human cardiovascular system, the proposed simulator consists of a left ventricle simulation module, an aorta simulation module, a peripheral resistance simulation module, and a positive/negative pressure control reservoir module. Simulating physiologies of blood pressure, the compliance chamber in the simulator can control arterial stiffness to produce age-dependent pressure waveforms. The augmentation index was used to assess the pressure waveforms generated by the simulator. The test results show that the simulator can generate and control radial pressure waveforms similar to human pulse signals consisting of early systolic pressure, late systolic pressure, and dicrotic notch. Furthermore, the simulator’s left ventricular pressure-volume loop results demonstrate that the simulator exhibits mechanical characteristics of the human cardiovascular system. The proposed device can be effectively used as a “standard” radial artery pressure simulator to calibrate the wearable sensor’s measurement characteristics and to develop more advanced sensors. The simulator is intended to serve as a platform for the development, performance verification, and calibration of wearable blood pressure sensors. It will contribute to the advancement of the wearable blood pressure sensor technology, which enables real-time monitoring of users’ radial artery pressure waveforms and eventually predicting cardiovascular diseases.

Funder

Korea Institute of Oriental Medicine

Publisher

Hindawi Limited

Subject

Health Informatics,Biomedical Engineering,Surgery,Biotechnology