Effect of Spherical Aberration on the Optical Quality after Implantation of Two Different Aspherical Intraocular Lenses

Author:

Lasta Michael12ORCID,Miháltz Kata12ORCID,Kovács Illés3,Vécsei-Marlovits Pia Veronika12

Affiliation:

1. Department of Ophthalmology, Hospital Hietzing, Vienna, Austria

2. Karl Landsteiner Institute of Process Optimization and QM in Cataract Surgery, Vienna, Austria

3. Department of Ophthalmology, Semmelweis University, Budapest, Hungary

Abstract

Purpose. To compare the effect of spherical aberration on optical quality in eyes with two different aspherical intraocular lenses. Methods. 120 eyes of 60 patients underwent phacoemulsification. In patients’ eyes, an aberration-free IOL (Aspira-aA; Human Optics) or an aberration-correcting aspherical IOL (Tecnis ZCB00; Abott Medical Optics) was randomly implanted. After surgery, contrast sensitivity and wavefront measurements as well as tilt and decentration measurements were performed. Results. Contrast sensitivity was significantly higher in eyes with Aspira lens under mesopic conditions with 12 cycles per degree (CPD) and under photopic conditions with 18 CPD (p=0.02). Wavefront measurements showed a higher total spherical aberration with a minimal pupil size of 4 mm in the Aspira group (0.05 ± 0.03) than in the Tecnis group (0.03 ± 0.02) (p=0.001). Strehl ratio was higher in eyes with Tecnis (0.28 ± 0.17) with a minimal pupil size larger than 5 mm than that with Aspira (0.16 ± 0.14) (p=0.04). In pupils with a minimum diameter of 4 mm spherical aberration had a significant effect on Strehl ratio, but not in pupils with a diameter less than 4 mm. Conclusions. Optical quality was better in eyes with the aberration-correcting Tecnis IOL when pupils were large. In contrast, this could not be shown in eyes with pupils under 4 mm or larger. This trial is registered with Clinicaltrials.gov NCT03224728.

Publisher

Hindawi Limited

Subject

Ophthalmology

Cited by 17 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3