Affiliation:
1. Risø National Laboratory, Roskilde, Denmark
Abstract
A new apparatus for dynamical texture measurements using neutron diffraction has been
developed. A variable wavelength neutron spectrometer has been modified to incorporate a linear
position-sensitive detector. The sample is orientated by a fully automatic Euler goniometer, and
it can be heated by a stream of hot air at rates up to 200℃/min. A quarter of a complete pole
figure can be recorded in 14 minutes with an accuracy, determined by counting statistics, of
2–3%. The experimental set-up is described in the present paper and the performance, including
sources of systematical errors, is discussed. Finally the viability of this apparatus is demonstrated
by studies of the recrystallization kinetics of 95% cold rolled copper.