Al2O3 Particles on Titanium Dental Implant Systems following Sandblasting and Acid-Etching Process

Author:

Schupbach Peter1ORCID,Glauser Roland2,Bauer Sebastian3

Affiliation:

1. Schupbach Ltd, Laboratory for Histology, Electron Microscopy and Imaging, CH-8800 Thalwil, Switzerland

2. Cosmodent Clinic, CH-8001 Zurich, Switzerland

3. Material Research and Surface Technologies, Nobel Biocare Services AG, CH-8302 Kloten, Switzerland

Abstract

Dental implants with moderately rough surfaces show enhanced osseointegration and faster bone healing compared with machined surfaces. The sandblasting and acid-etching (SA) process is one technique to create moderately rough dental implant surfaces. The purpose of this study was to analyse different commercially available implant systems with a SA-modified surface and to explore the widespread notion that they have similar surface properties regarding morphology and cleanliness. SA-modified surfaces of nine implant systems manufactured by Alpha-Bio Tec Ltd, Camlog Biotechnologies AG, Dentsply Sirona Dental GmbH, Neoss Ltd, Osstem Implant Co. Ltd, Institute Straumann AG, and Thommen Medical AG were analyzed using scanning electron microscopy (SEM) and energy dispersive X-ray spectroscopy (EDX) and examined for surface cleanliness. Six implants from three different lots were selected per each implant system. Mean particle counts for each implant and the mean size of the particles were calculated from three different regions of interest and compared using ANOVA and Tukey’s test. SEM analysis showed presence of particles on the majority of analyzed implant surfaces, and EDX evaluations determined that the particles were made of Al2O3 and thus remnants of the blasting process. SPI®ELEMENT INICELL® and Bone Level (BL) Roxolid® SLActive® implant surfaces showed the highest mean particle counts, 46.6 and 50.3 per area, respectively. The surface of BL Roxolid® SLActive® implant also showed the highest variations in the particle counts, even in samples from the same lot. The mean size of particles was 1120±1011 μm2, measured for USIII CA Fixture implants, while the biggest particle was 5900 μm2 found on a BL Roxolid® SLActive® implant. These results suggest that not all manufacturers are able to produce implant surfaces without particle contamination and highlight that the surface modification process with the SA technique should be appropriately designed and controlled to achieve a clean and consistent final medical device.

Publisher

Hindawi Limited

Subject

Biomedical Engineering,Biomaterials

Cited by 36 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3