Usefulness of 99mTc-Pertechnetate SPECT-CT in Thyroid Tissue Volumetry: Phantom Studies and a Clinical Case Series

Author:

Calandri Enrico1,Giraudo Maria Teresa2,Sirovich Roberta2,Ostan Antonella3,Pultrone Mirco1,Frantellizzi Viviana4,Conversano Lucia1,Bagnasacco Paolo5,Gallina Sonya1,Vincentis Giuseppe De6

Affiliation:

1. Department of Medicine and Urgency, Nuclear Medicine Unit, Ospedale degli Infermi, Via dei Ponderanesi 2, Ponderano (BI), Italy

2. Department of Mathematics G. Peano, University of Torino, Torino, Italy

3. Department of Health Physics, Azienda Ospedaliera Maggiore della Carità, Novara, Italy

4. Department of Radiological Sciences, Oncology and Anatomical Pathology, Nuclear Medicine Unit, “Sapienza” University, Rome, Italy

5. Department of Medicine and Urgency, Radiotherapy Unit, Ospedale degli Infermi, Ponderano (BI), Italy

6. Department of Radiological Sciences, Oncology and Anatomical Pathology, Nuclear Medicine Unit, “Sapienza” University, Rome, Italy.

Abstract

Background: An accurate measurement of the target volume is of primary importance in theragnostics of hyperthyroidism. Objective: Our purpose was to evaluate the accuracy of a threshold-based isocontour extraction procedure for thyroid tissue volumetry from SPECT-CT. Methods: Cylindrical vials with a fixed volume of 99mTcO4 at different activities were inserted into a neck phantom in two different thickness settings. Images were acquired by orienting the phantom in different positions, i.e., 40 planar images and 40 SPECT-CT. The fixed values of the isocontouring threshold for SPECT and SPECT-CT were calculated by means of linear and spline regression models. Mean, Median, Standard Deviation, Standard Error, Mean Absolute Percentage Error and Root Mean-Square Error were computed. Any difference between the planar method, SPECT and SPECT-CT and the effective volume was evaluated by means of ANOVA and posthoc tests. Moreover, planar and SPECT-CT acquisitions were performed in 8 patients with hyperthyroidism, considering relevant percentage differences greater than > 20% from the CT gold standard. Results: Concerning phantom studies, the planar method shows higher values of each parameter than the other two methods. SPECT-CT shows lower variability. However, no significant differences were observed between SPECT and SPECT-CT measurements. In patients, relevant differences were found in 7 out of 9 lesions with the planar method, in 6 lesions with SPECT, but in only one with SPECT-CT. Conclution: Our study confirms the superiority of SPECT in volume measurement if compared with the planar method. A more accurate measurement can be obtained from SPECT-CT.

Publisher

Bentham Science Publishers Ltd.

Subject

Pharmacology,Radiology, Nuclear Medicine and imaging

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3