Author:
Cervino Gabriele,Romeo Umberto,Lauritano Floriana,Bramanti Ennio,Fiorillo Luca,D’Amico Cesare,Milone Dario,Laino Luigi,Campolongo Francesco,Rapisarda Silvia,Cicciù Marco
Abstract
Purpose:
The objective of this investigation is to study prosthodontics and internal components resistance to the masticatory stress and considering different force directions by using Finite Element Method analysis (FEM). The structural materials of the components are usually Titanium alloy grade 4 or 5 and thus, guarantee the integration of the fixture in the bone due to the osteointegration phenomena. Even if the long-term dental implant survival rate is easy to be obtained and confirmed by numerous researches, the related clinical success, due to the alteration of the mechanical and prosthodontics components is still controversial.
Methods:
By applying engineering systems of investigations like FEM and Von Mises analyses, it has been investigated how dental implant material was held against the masticatory strength during the dynamic masticatory cycles. A three-dimensional system involved fixture, abutment and the connection screws, which were created and analyzed. The elastic features of the materials used in the study were taken from recent literature data.
Results:
Data revealed a different response for both types of devices, although implant neck and dental abutment showed better results for all conditions of loading while the abutment screw represented aweak point of the system.
Conclusion:
The data of this virtual model showed all the features of different prosthetic retention systems under the masticatory load. Clinicians should find better prosthetic balance in order to better distribute the stress over the component and to guarantee patients’ clinical long-term results.
Publisher
Bentham Science Publishers Ltd.
Reference35 articles.
1. Bramanti E, Cervino G, Lauritano F, et al.
FEM and von mises analysis on prosthetic crowns structural elements: Evaluation of different applied materials.
Sci World J
2017;
1029574.
2. van Staden RC, Guan H, Johnson NW, Loo YC, Meredith N.
Step-wise analysis of the dental implant insertion process using the finite element technique.
Clin Oral Implants Res
2008;
19
(3)
: 303-13.
3. Kitagawa T, Tanimoto Y, Nemoto K, Aida M.
Influence of cortical bone quality on stress distribution in bone around dental implant.
Dent Mater J
2005;
24
(2)
: 219-24.
4. Cicciu M, Bramanti E, Matacena G, Guglielmino E, Risitano G.
FEM evaluation of cemented-retained versus screw-retained dental implant single-tooth crown prosthesis.
Int J Clin Exp Med
2014;
7
(4)
: 817-25.
5. Desai SR, Karthikeyan I, Gaddale R.
3D finite element analysis of immediate loading of single wide versus double implants for replacing mandibular molar.
J Indian Soc Periodontol
2013;
17
(6)
: 777-83.
Cited by
81 articles.
订阅此论文施引文献