Abstract

Changes to the bearing surfaces of eighteen uncemented total hip replacements retrieved at revision surgery were assessed by three-dimensional binocular microscopy, Rank Taylor Hobson talysurf measurements, scanning electron microscopy, and X-ray dispersive analysis. Abrasions on the non-articular surface of the polyethylene cups were present. Bone particles were found in tracks in the bearing surfaces of both the titanium femoral heads and the polyethylene cups and were responsible for wear of these surfaces. Although the wear of the femoral heads appeared substantial to naked eye examination, the surface finish of these surfaces remained within the British ISO standards for titanium alloy when assessed by the methods used above. This study concluded that direct contact between polyethylene and bone should be avoided in total hip arthroplasty and that ‘third body’ wear from bone particles occurred in these uncemented prostheses. Both components of this type of implant should be replaced at revision surgery and titanium should be avoided as a bearing surface in hip arthroplasty. Present methods of assessing the surface finish of titanium should be re-evaluated and more reliable ones considered.

Publisher

SAGE Publications

Subject

Mechanical Engineering,General Medicine

Cited by 5 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Methods for Surface Superfinishing of Prosthesis;Advanced Micro- and Nano-manufacturing Technologies;2021-10-02

2. Modelling of loss mechanisms in a pushing metal V-belt continuously variable transmission. Part 1: Torque losses due to band friction;Proceedings of the Institution of Mechanical Engineers, Part D: Journal of Automobile Engineering;2004-11-01

3. The influence of femoral head surface roughness on the wear of ultrahigh molecular weight polyethylene sockets in cementless total hip replacement;Journal of Biomedical Materials Research;1999

4. Effect of Counterface Material on the Characteristics of Retrieved Uncemented Cobalt-Chromium and Titanium Alloy Total Hip Replacements;Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine;1996-09

5. Estimation of Wear in Total Hip Replacement Using a Ten Station Hip Simulator;Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine;1996-09

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3