Affiliation:
1. DePuy International Limited, Leeds, UK
2. School of Mechanical Engineering, University of Leeds, Leeds, UK
Abstract
The possibility of replacing the femoral head alone, in either solid or articular surface replacement form, during revision operations on metal-on-metal total hip replacements remains an attractive feature of such implants. In the present investigation, laboratory simulator studies of the influence upon volumetric wear of inserting a new femoral head, of introducing some head rotation, and of damaging the femoral head by scratches have all been explored. New and rotated heads both involve an additional running-in period, but the experimental studies show that the volumetric wear associated with this process is less than the initial running-in wear. The beneficial effects upon volumetric wear of small clearances have been confirmed, while the processing of high-carbon Co-Cr-Mo materials appears to be much less influential. Scratches did not affect wear as much as head replacement or head rotation, but the ongoing wear rates were somewhat higher.
Subject
Mechanical Engineering,General Medicine
Cited by
15 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献