Affiliation:
1. Department of Orthopaedic Surgery and Rehabilitation, University of Nebraska Medical Center, Omaha, Nebraska, USA
Abstract
Arthroplasty outcome is influenced by the ‘quality’ of bone preparation for implant insertion. Surgeons face increasing choices of technique and instrumentation, yet clinical scoring methods assess the overall outcome and patient satisfaction but not the bone cuts directly. ‘Quality management’ of bone reshaping is needed to evaluate different bone cutting methods and computer assisted orthopaedic surgery (CAOS) systems. Analyses and experiments in this study were formulated for measurement and computation of four quantitative characteristics of bone preparation ‘quality’ and produced a highly condensed index for each. These represented (a) surface finish of cuts, (b) implant fit/looseness possible with the cut shape, (c) implant location/misalignment, and (d) accuracy of individual planar cuts. Assessment of synthetic bone cuts verified the robustness of the method for wide application in arthroplasty intraoperatively, in vitro and for comparing navigation systems.
Subject
Mechanical Engineering,General Medicine
Cited by
12 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献