A novel esophageal retractor with eccentric balloon during atrial fibrillation ablation

Author:

Dai Wen‐Li1,Yao Ke‐Xin1,Li Meng‐Meng1ORCID,Li Song‐Nan1,Sang Cai‐Hua1ORCID,Jiang Chen‐Xi1ORCID,Guo Xue‐Yuan1,Li Xu1,Feng Li1,Jia Chang‐Qi1,Ning Man1,Dong Jian‐Zeng1ORCID,Ma Chang‐Sheng1ORCID

Affiliation:

1. Department of Cardiology National Clinical Research Center for Cardiovascular Diseases Beijing Anzhen Hospital Capital Medical University Chaoyang‐qu Beijing China

Abstract

AbstractBackgroundDue to the anatomically adjacent relationship between the left atrium (LA) and esophagus, energy delivery on the posterior wall of LA is limited. The aim of this study was to evaluate the feasibility of a novel esophageal retractor (SAFER) with an inflatable C‐curve balloon during atrial fibrillation (AF) ablation.MethodNine patients underwent AF ablation assisted with the SAFER. After inflation, the esophagus was deviated laterally away from the intended ablation site of the posterior wall under local anesthesia. The extent of mechanical esophageal deviation (MED) was evaluated under fluoroscopy, defined as the shortest distance from the trailing esophageal edge to the closest point of the ablation line. Gastroscopy was performed before and after ablation. The target ablation index used in all LA sites including the posterior wall was 400–450 after effective MED. All adverse events during the periprocedural period were recorded.ResultsThe mean deviation distance achieved 16.2 ± 9.6 mm away from the closest ablation point of the pulmonary vein lesion set. With respect to the individual left and right pulmonary vein lesion sets, the deviation distance was 19.7 ± 11.5 and 12.7 ± 6.8 mm, respectively. The extent of deviation was 0 to 5 mm, 5.1 to 10 mm, or >10 mm in 0(0%), 7(38.9%), and 11(61.1%), respectively. Procedural success was achieved in all patients without acute reconnection. There was only one esophageal complication which manifested as esophageal erosion and this patient experienced throat pain possibly related to the SAFER retractor with no clinical sequelae.ConclusionEsophageal deviation with the novel eccentric balloon is a novel feasible choice during AF ablation, enabling adequate energy delivery to the posterior wall of LA. Additional prospective randomized controlled studies are required for further validation.

Funder

National Key Research and Development Program of China

Publisher

Wiley

Subject

Cardiology and Cardiovascular Medicine,General Medicine

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Esophageal-Pericardial and Esophageal-Atrial Fistulae;Thoracic Surgery Clinics;2024-06