Influence of pulse energy, tip design and insertion depth during Er:YAG‐activated irrigation on cleaning efficacy in simulated severely curved complex root canal systems in vitro

Author:

Bürklein Sebastian1ORCID,Abdi Izzeddin1,Schäfer Edgar1ORCID,Appel Carsten2,Donnermeyer David3ORCID

Affiliation:

1. Central Interdisciplinary Ambulance in the School of Dentistry University of Münster Münster Germany

2. Private Practice, Endodontie am Venusberg Bonn Germany

3. Department of Periodontology and Operative Dentistry University of Münster Münster Germany

Abstract

AbstractAimTo investigate the influence of pulse energy, tip geometry and tip position in simulated 3D‐printed root canals with multiple side canals at different levels in all directions on the cleaning performance of laser‐activated irrigation (LAI) compared to sonic activation (EDDY) and conventional needle irrigation (NI).Methodology3D‐printed root canal models (25/.06, length 20 mm, curvature 60°, radius 5 mm) with side canals (diameter 0.2 mm) at 2, 5 and 8 mm from the apex were filled with coloured biofilm‐mimicking hydrogel. LAI (Morita AdvErL Evo, Kyoto, Japan) was performed with six settings (n = 20; pulse‐energy, pulses per second [PPS], tip position): LAI1 (50 mJ, 25 PPS, P400FL, canal entrance [CE]), LAI2 (same as LAI1, but insertion depth 9 mm before the apical endpoint [AE] [corresponding to 1 mm above the first lateral canals]), LAI3 (80 mJ, 25 PPS, P400FL, 9 mm before AE), LAI4 (same as LAI 3, but at CE) for 3 × 20 s each, LAI5 (50 mJ, 25 PPS, P400FL 2 × 20 s, CE & R200T (30 mJ, 25 PPS, 1 × 20 s, 9 mm before AE), LAI6 (30 mJ, 25 PPS, R200T, 9 mm before AE, 3 × 20 s). A continuous irrigation (3 mL/20 s) using distilled water accompanied the irrigation cycles. NI and EDDY (3 × 20 s each; 3 mL/20 s irrigation, insertion AE minus 1 mm, amplitude 4 mm) served as control groups. Biofilm‐mimicking hydrogel removal (ImageJ, NIH) was assessed for the entire system, the central canal and the lateral canals using standardized photographs with a microscope (Expert DN, Müller‐Optronic) and statistically analysed was performed using Kruskal–Wallis and Dunn tests (p = .05). Irrigant extrusion beyond the foramina was also recorded.ResultsLAI2 (99.08%; interquartile range [IQR]: 96.85–100.00) and LAI3 (97.50%; 96.24–100.00) achieved the significantly best and LAI6 (80.08%; 73.41–84.69) the significantly worst removal of hydrogel from the entire root canal system amongst all LAI configurations (p < .05). There were no significant differences between LAI6, EDDY (72.89%; 67.49–76.22) and manual irrigation (54.39%; 51.01–56.94) (p > .05). R200T laser tip caused significantly more often irrigant extrusion than all other techniques (p < .05).ConclusionTip design, energy settings, and the positioning of the laser tip below the canal entrance caused an improvement in cleaning performance of the LAI. However, the small R200T tip created significantly more procedural errors (irrigant extrusion) due to higher concentrated energy.

Publisher

Wiley

Subject

General Dentistry

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3