The aerodynamic particle-size distribution for two doses of a Becloforte metered-dose inhaler (MDI) was measured by use of a twin-stage impinger (TSI), the new multi-stage (five-stage) liquid impinger (MSLI) and the Andersen cascade impactor (ACI) (n = 5 for each apparatus). The mean (s.d.) fine-particle doses measured by the three techniques for the Becloforte MDI were 40.3 (1.2), 45.7 (0.5) and 41.8 (0–4)% w/w, respectively; the median mass aerodynamic diameters (MMAD) measured using the MSLI and the ACI were 3.50 and 3.73 μm, respectively. The MSLI fine particle (< 6.8 μm) doses for 2, 5, 10, 20, 30 and 40 doses from Becloforte MDIs (n = 5 for each dose) were 49.7 (0.7), 52.9 (1.2), 45.3 (0.6), 45.5 (0.71), 45.9 (0.7) and 46.4 (0.7)% w/w, respectively. Values obtained using the ACI (< 5.8 μm) were 40.8 (1.0), 41.0 (0.8), 44.4 (0.5), 43.1 (0.4), 42.8 (0.5) and 40.4 (0.4)% w/w (n = 4). MMAD values measured with the MSLI were 3.39, 3.46, 3.75, 3.91, 4.15 and 4.45 μm, respectively; using the ACI they were 3.46, 3.54, 3.61, 3.66, 3.73 and 3.85 μm.
The results indicate that the measured aerodynamic particle-size distributions of beclomethasone dipropionate MDIs are affected by the dose dispensed and by the apparatus used for measurement.
Publisher
Oxford University Press (OUP)
Subject
Pharmaceutical Science,Pharmacology
Reference16 articles.
1. Inhalations: collaborative study on the measurement of the fine-particle dose using inertial impactor;Aiche;Pharmeuropa,1993
2. Comparisons of aerodynamic particle size from 250 μg per dose beclomethasone dipropionate metered-dose inhalers;Barnes;Int. J. Pharm.,1996
3. Recommendations of the USP advisory panel on aerosols on the USP general chapters on aerosols [601] and uniformity of dosage units [905];Byron;Pharm. Forum.,1994
4. Low first spray drug content in albuterol metered-dose inhalers;Cyr;Pharm. Res.,1991
5. The influence of sampling chamber dimensions on aerosol particle size measurements by cascade impactor and twin impinger;Fults;J. Pharm. Pharmacol.,1991
Cited by
7 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献