Limited Revision with a Newly Designed Hinge Device for the Treatment of Mega‐Prosthesis Hinge Failure: Two Case Reports

Author:

Zheng Kai1ORCID,Yu Xiu‐chun1,Xu Ming1ORCID,Cui Hao‐cheng1,Zhang Jing‐yu2ORCID,Wei Yue3

Affiliation:

1. Department of Orthopedics The 960th Hospital of the PLA Joint Logistice Support Force Jinan China

2. Department of Bone Tumor Tianjin Hospital Tianjin China

3. Shandong Weigao Orthopaedic Device Co., Ltd. Weihai China

Abstract

BackgroundSurgical treatment for hinge failure in mega‐prosthesis continues to be a challenge. This study introduces a new method for treating hinge failure by using a unilateral prosthesis and hinge revision.Case PresentationWe here present two patients who underwent mega‐prosthesis reconstruction after resection of osteosarcoma in the distal femur. To address the issue of knee hyperextension after mega‐prosthesis reconstruction, one patient underwent three revision surgeries, two surgeries were performed using the original hinge, and one surgery involved a newly designed hinge. To resolve the problem of dislocation, one patient underwent three revisions, with the first two revisions not involving hinge replacement and the third revision involving a newly designed hinge. Two replacements of unilateral prosthesis and hinge renovations were successful.ConclusionsUnilateral prosthesis and newly designed hinge device revision are effective in treating the failure of old‐fashioned mega‐prosthesis hinges.

Publisher

Wiley

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3