Abstract
The method is presented for converting a coherent population trapping (CPT) atomic clock into a CPT vector (compass-) magnetometer without mechanically moving parts through a relatively simple add-on. The system of 3D Helmholtz coils was used for compensation of the external magnetic field, allowing measurement of both strength and direction of a magnetic field at the sensitivity level of
s
u
b
−
n
T
/
H
z
1
/
2
within a 10-Hz bandwidth. Angular resolution of the developed vector magnetometer amounts to about
1
0
−
2
degrees.