Affiliation:
1. Ghent University Centre for X-Ray Tomography (UGCT)
2. Paul Scherrer Institute
3. Institute for Biomedical Engineering
Abstract
X-ray dual-phase grating interferometry provides quantitative micro-structural information beyond the optical resolution through its tunable correlation length. Ensuring optimal performance of the set-up requires accurate correlation length estimation and precise alignment of the gratings. This paper presents an automated procedure for determining the complete geometrical parameters of the interferometer set-up with a high degree of precision. The algorithm’s effectiveness is then evaluated through a series of experimental tests, illustrating its accuracy and robustness.
Funder
Interreg Vlaanderen-Nederland
European Regional Development Funds
Provincie Oost-Vlaanderen
Fonds Wetenschappelijk Onderzoek
Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung
Personalized Health and Related Technologies (PHRT) (