Affiliation:
1. University of Rochester
Abstract
A co-registered speckle contrast optical tomography and frequency domain-diffuse optical tomography system has been designed for imaging total hemoglobin concentration, blood oxygenation, and blood flow with the future aim of monitoring Jones fractures of the fifth metatarsal. Experimental validation was performed using both in vitro tissue-mimicking phantoms and in vivo cuff occlusion experiments. Results of these tissue phantom experiments ensure accurate recovery of three-dimensional distributions of optical properties and flow. Finally, cuff occlusion experiments performed on one healthy human subject demonstrate the system’s ability to recover both decreasing tissue oxygenation and blood flow as caused by an arterial occlusion.
Funder
National Institute of Arthritis and Musculoskeletal and Skin Diseases
Subject
Atomic and Molecular Physics, and Optics,Biotechnology
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献