Abstract
The technique of remote refocusing is used in optical microscopy to
provide rapid axial scanning without mechanically perturbing the
sample and in techniques such as oblique plane microscopy that build
on remote refocusing to image a tilted plane within the sample. The
magnification between the pupils of the primary (O1) and secondary
(O2) microscope objectives of the remote-refocusing system has been
shown previously by Mohanan and Corbett [J. Microsc.
288, 95 (2022)JMICAR0022-272010.1111/jmi.12991] to be
crucial in obtaining the broadest possible remote-refocusing range. In
this work, we performed an initial alignment of a remote-refocusing
system and then studied the effect of axial misalignments of O1 and
O2, axial misalignment of the primary tube lens (TL1) relative to the
secondary tube lens (TL2), lateral misalignments of TL2, and changes
in the focal length of TL2. For each instance of the setup, we
measured the mean point spread function FWHMxy of 100 nm fluorescent beads and the
normalized bead integrated fluorescence signal, and we calculated the
axial and lateral distortion of the system; all of these quantities
were mapped over the remote-refocusing range and as a function of
lateral image position. This allowed us to estimate the volume over
which diffraction-limited performance is achieved and how this changes
with the alignment of the system.