Higher Wear-Rate of Third-Generation Metal-Backed Reflection Cups with Eto-Sterilised UHMWPE at a Mean 13 Years Follow-up

Author:

Hengst David M.1,Thomsen Per B.1,Homilius Morten1,Hansen Torben B.1,Stilling Maiken1

Affiliation:

1. Department of Orthopaedic Surgery, Hospitalsenheden Vest, Holstebro, Denmark University of Aarhus, Aarhus - Denmark

Abstract

Polyethylene (PE) wear and osteolysis is a recognised problem with non-cross linked PE liners and first generation modular cup designs. Wear particles induce osteolysis leading to aseptic loosening. We retrospectively compared the linear PE wear and implant survival and revision rates of the Reflection Cup and the Duraloc 300. After a mean clinical follow-up of 13 years (range 11-15 years), the 2D linear PE wear-rate of the Reflection liner (n = 68) was 0.23 mm/year, with a mean total wear of 3.14 mm (1.04-7.36), SD 1.45. The wear-rate of the Duraloc 300 cups (n = 32) was 0.14 mm/year, with a mean total wear of 1.84 mm (0.55-4.63), SD 1.07. The difference in PE wear-rate as well as mean total wear was highly significant (p = 0.0001). There was a positive correlation between wear-rate and both Oxford Hip Score and Harris Hip Score (p = 0.02). Large acetabular cup size (>54 mm), HA coating on the stem and age <50 years did not influence PE wear. The higher wear-rate in the Reflection liners could be related to the EtO sterilisation. Intermediate and long-term follow-up is advisable.

Publisher

SAGE Publications

Subject

Orthopedics and Sports Medicine,Surgery

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3