Catheter Fracture and Embolization Associated with Arm Implantation of the Cook Vital Port

Author:

Burbridge Brent1,Stoneham Grant1,Szkup Peter1,Otani Rob1,Kriegler Stefan1

Affiliation:

1. Medical Imaging, Royal University Hospital, Saskatoon, Saskatchewan, Canada

Abstract

Purpose To determine the frequency of arm port catheter fracture and embolization related to the Cook Vital Port Mini Titanium. Materials and Methods A retrospective audit of our Cerner Radiology Information System was performed between June 1, 2006, and June 30, 2011, to determine the number of Cook arm venous ports implanted and the frequency of foreign body retrievals related to catheter fracture for these arm ports. Results A total of 691 arm implantations of the Cook Vital Port during the 5-year time frame were analysed. Eleven of these patients (1.6%) required intravenous foreign body retrieval in the interventional radiology suite related to catheter fracture and embolization. Three of these fractured catheters were retrieved from the peripheral venous system upstream of the pulmonary circulation, whereas 8 embolized to the pulmonary arteries. All were successfully extracted with an intravenous snare by interventional radiology. Conclusion We discovered a 1.6% frequency of catheter fracture and embolization associated with arm implantation of the Cook Vital Port. All the catheters fractured at the vein entry site and did not detach from the port housing. The cause for catheter fracture and embolization is uncertain. Pulmonary embolization of the fractured catheters puts the patients at risk for possible further complications. No patients had ancillary complications related to catheter embolization or to catheter extraction procedures. Further investigation is required in an attempt to determine the circumstances that may result in catheter fracture and embolization related to this venous access device.

Publisher

SAGE Publications

Subject

Radiology, Nuclear Medicine and imaging,General Medicine

Reference16 articles.

1. Is radiologic placement of an arm port mandatory in oncology patients?

2. Long-term outcomes of peripheral arm ports implanted in patients with colorectal cancer

3. RaffJ.M. HeilD. RollinsD.L. Vascular access, pumps, and infusion McKennaR.J. MurphyG.P. Cancer Surgery 1994 Lippincott, Williams & Wilkins Philadelphia 47–62.

Cited by 13 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献