Affiliation:
1. Department of Chemistry, California State University-Fresno, Fresno, California 93740-0070
Abstract
A micronebulizer/tangential torch configuration was used for the direct introduction of organic and aqueous aerosols into a 120-W argon Beenakker cavity plasma. Atomic emission was measured for Ca, Mg, Cr, Zn, Li, and Sr. Detection limits (3σ) with the use of aqueous standards were 40, 80, 180, 535, 9, and 90 ng/mL, respectively. Detection limits under similar conditions with the use of 50% methanol-aqueous solutions were 35, 107, 230, 10, and 127 ng/mL for Ca, Mg, Cr, Li, and Sr, respectively. Iodine detection in the helium plasma was evaluated in aqueous, 50% methanol-aqueous, and pure methanol solutions, with detection limits of 17, 25, and 42 μg/mL, respectively.
Subject
Spectroscopy,Instrumentation
Cited by
13 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献