Investigation of Electrical and Optical Subtractions Using Two Input-Port and Two Output-Port FT-IR Spectrometers

Author:

Laroche Gaétan1,Giroux Jean1,Bordeleau Alain1,Garneau Jean-Marc1

Affiliation:

1. Defense Research Establishment, Valcartier, 2459, boul Pie XI, Nord, C.P. 8800, Courcelette, Québec GOA 1R0, Canada (G.L., J.-M.G.); and Bomem Inc. (Hartmann & Braun), 450 St-Jean Baptiste, Québec, Québec G2E 5S5, Canada (J.G., A.B.)

Abstract

Two FT-IR spectrometers, each using two input ports and two output ports, have been used to minimize the effect of background noise and source fluctuation noise in infrared emission spectra of various sources. Blackbody sources, propane/air flames, and infrared flares have been studied, and spectra were recorded in the spectral region ranging from 1.7 to 5 μm. With the use of the two input-port and one output-port configurations, it was found that real-time optical subtraction could generate 80% background-noise-free spectra. When the spectrometers were operated in the one input-port and two output-port configurations, spectra that were free of source fluctuation noise were obtained with the use of real-time electrical subtraction of signals measured at both detectors. New signal processing techniques have thus been developed. An increase in the signal-to-fluctuation-noise ratio by a factor of seven has been observed in the interferograms, which in turn leads to a 2 × increase of the signal-to-noise ratio in the corresponding spectra. During this signal processing sequence requiring the use of two analog-to-digital converters (ADC) (one for each detector channel), intensity information was then lost, so that no calibrated spectra could be measured. However, with the use of a single-channel ADC, it was shown that, by a process of simply subtracting signals recorded from both detectors operated under similar amplifier gain, fluctuation noise could be partly removed and intensity information could also be retained. In conjunction with the high scanning velocity of the interferometer (60 scans/s at a 16-cm−1 resolution), this technique has proven to be very useful in measuring emission spectra of highly fluctuating infrared sources, such as flares.

Publisher

SAGE Publications

Subject

Spectroscopy,Instrumentation

Cited by 4 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3