Grating based X-ray phase contrast CT imaging with iterative reconstruction algorithm

Author:

Qi Jun-Cheng ,Chen Rong-Chang ,Liu Bin ,Chen Ping ,Du Guo-Hao ,Xiao Ti-Qiao , ,

Abstract

Grating based X-ray imaging technology is a coherent imaging technique that bears tremendous potential in three-dimensional tomographic imaging of weak absorption contrast specimens. Three kinds of contrast information including absorption, phase and scattering can be retrieved separately based on a single set of raw projections. However, the grating based X-ray imaging with the conventional phase-retrieval method using the conventional phase-stepping approach and filtered back projection (FBP) reconstruction algorithm require large amounts of raw data, so that long exposure time and large amounts of radiation dose is accepted by the sample. According to the traditional grating based X-ray imaging system, we propose a low dose, fast, multi-contrast CT reconstruction approach based on the iterative reconstruction algorithm that optimizes dose efficiency but does not share the main limitations of other reported methods. Prior to reconstruction, firstly, the projections are acquired with the phase stepping approach and multi-contrast projections are retrieved from the raw data by conventional retrieval algorithm. Then the rotational variable differential phase projections are converted to rotational invariable projections by means of decomposing the differential phase projections into the rotational invariable projections in two mutually perpendicular derivative directions via the transformation of coordinates. Finally, the absorption, phase and scattering information are reconstructed with the iterative reconstruction algorithm and the phase is retrieved based on the fast Fourier transform (FFT). We validated and assessed the phase reconstruction approach with a numerical simulation on a phase Shepp-Logan phantom. The experiment was performed at the X-ray imaging and biomedical application beam line (BL-13W) in the Shanghai Synchrotron Radiation Facility (SSRF) where 20 keV X-ray from a Si(111) monochromator is emitted. The X-ray interferometer was positioned at 34 m from the Wiggler source. The images were recorded with a scintillator/lens-coupled CCD camera with 2048 pixel2048 pixel resolution and an effective pixel size of 9 m. The numerical tests and the experimental results demonstrate that, for the small radiation dose deposited in the sample, the iterative reconstruction algorithm provides phase reconstructions of better quality and higher signal to noise ratio than the conventional FBP reconstruction algorithm, and also provides the multi-contrast 3D images, including absorption image, phase image and scattering image. This development is of particular interest for applications where the samples need inspecting under low dose and high speed conditions, and will play an important role in the nondestructive and quantitative imaging in the industry, biomedical and medical diagnosis fields.

Publisher

Acta Physica Sinica, Chinese Physical Society and Institute of Physics, Chinese Academy of Sciences

Subject

General Physics and Astronomy

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3