First evaluation of a novel ionisation chamber for thermal neutron beam monitoring

Author:

Maulerová VendulaORCID,Kanaki KalliopiORCID,Klein Martin,Schmidt Christian J.ORCID,Kadletz Peter M.ORCID,Laloni Alessio,Hall-Wilton Richard J.ORCID

Abstract

AbstractThe European Spallation Source ERIC (ESS), currently under construction in Lund, Sweden is a facility established to deliver the highest integrated neutron flux originating from a pulsed source with the aim of supporting an initial fifteen neutron instruments for cutting edge science experiments. This in turn requires reliable monitoring at complex neutron beam lines: in particular, linearity, timing capability, adaptability of the design for various flux ranges (dynamic range) and sensitivity to neutrons within the range of 0.6-10Å are expected from the neutron beam monitors to be installed at the ESS beam lines. Additionally, operational stability and low attenuation are also desirable characteristics for such neutron beam monitoring. A prototype neutron beam monitor based on the ionisation chamber principle and a boron converter, designed by CDT CASCADE Detector Technologies GmbH and ESS, has been investigated at the BER-II research reactor of Helmholtz Zentrum Berlin (HZB). The effort to design and investigate a thermal neutron ionisation beam monitor was initiated by adapting the concept of ionisation chambers previously known elsewhere. So far all the characterised neutron beam monitors discriminate neutron hits on a discrete event basis (pulse mode), whereas the beam monitor prototype introduced in this paper estimates the total flux as a function of current (current mode). While most other neutron beam monitoring devices and detectors rely upon a signal amplifying gain stage, the ionisation chamber operates without any gain and is consequently robust against typical detector ageing effects that compromise the sensitivity over time. The initial tests were performed at the ESS V20 test beam line under realistic conditions resembling those of the future pulses of ESS. The linearity is demonstrated for 3Å pulses in the flux range of 2-3 × 105 n/s/cm2 and for white pulses (0.6-10Å) in the range of 1-5 × 106 n/s/cm2. The timing behaviour resembles the data previously recorded at the V20 beam lines. This novel implementation of a neutron sensitive ionisation chamber shows great promise for beam monitoring and diagnostics at ESS. As the ionisation beam monitor itself is an entirely passive device, it is adequately robust to be employed in areas of high irradiation where no regular servicing or maintenance can be provided.

Publisher

Springer Science and Business Media LLC

Subject

Instrumentation

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3