Five-day Wearable Respiratory Support With a Novel Ambulatory Pulmonary Assist System in an Awake Ovine Model

Author:

Hong Yeahwa123,Shin Suji1,Nasim Umar1,Scala Helen M.1,Roberts Kalliope G.1,Potchernikov Alexander S.1,Woolley Brian E.1,Skoog David J.4,Bacchetta Matthew5,Cook Keith E.1

Affiliation:

1. Department of Biomedical Engineering, Carnegie Mellon University, Pittsburgh, PA.

2. Department of Surgery, University of Pittsburgh Medical Center, Pittsburgh, PA.

3. Department of Cardiothoracic Surgery, University of Pittsburgh Medical Center, Pittsburgh, PA.

4. Advanced Respiratory Technologies Inc, Pittsburgh, PA.

5. Department of Cardiac Surgery, Vanderbilt University Medical Center, Nashville, TN.

Abstract

Background. The pulmonary assist system (PAS) is a wearable respiratory support system that is currently under development for patients with chronic lung disease as a bridge to lung transplantation or as destination therapy. This study evaluates the long-term performance and biocompatibility of the PAS in a 5-d awake, ovine model. Methods. The PAS was attached to normal sheep in venovenous configuration. Components of the PAS included a 0.9 m2 surface area oxygenator and a lightweight, battery-powered axial flow pump. The system was also tested using the Abbott PediMag as the control pump. Each sheep was supported on the PAS for 5 d with 2 L/min blood flow and 4 L/min sweep gas. Activated clotting times of 200–240 s were maintained using intravenous heparin. Pump performance, oxygen transfer, oxygenator resistance, and hematologic parameters were measured throughout the support. Results. The PAS, either using the axial flow pump or PediMag (n = 4 each), was well tolerated by the sheep without signs of device-related organ damage or hemolysis. All the studies achieved the full, 5-d study duration. The oxygenator resistance remained consistent without significant clot formation in all experiments with an average resistance of 2.55 ± 0.10 mm Hg/(L/min). The system achieved an average oxygen transfer rate of 116.4 ± 5.5 mL/min, with an average Hb concentration of 9.2 ± 0.6 g/dL. White blood cell, platelet, and hematocrit levels also remained stable and within normal limits throughout the study period. Conclusions. The PAS provided 5 d of uncomplicated ambulatory respiratory support with minimal clot formation, stable gas exchange, blood flow resistance, and hematologic parameters.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Reference32 articles.

1. Mortality in the United States, 2019.;Kochanek;NCHS Data Brief,2020

2. Exacerbations of chronic obstructive pulmonary disease.;Anzueto;Proc Am Thorac Soc,2007

3. Ventilator-induced lung injury [published correction appears in N Engl J Med;Slutsky;N Engl J Med,2013

4. Ventilator-induced lung injury: from the bench to the bedside.;Tremblay;Intensive Care Med,2006

5. Extracorporeal membrane oxygenation with spontaneous breathing as a bridge to lung transplantation.;Nosotti;Interact Cardiovasc Thorac Surg,2013