Influence of Tracheal Obstruction on the Efficacy of Superimposed High-frequency Jet Ventilation and Single-frequency Jet Ventilation

Author:

Sütterlin Robert1,LoMauro Antonella1,Gandolfi Stefano1,Priori Rita1,Aliverti Andrea1,Frykholm Peter1,Larsson Anders1

Affiliation:

1. From the Hedenstierna Laboratory, Anaesthesiology and Intensive Care, Department of Surgical Sciences, Uppsala University, Uppsala, Sweden (R.S., P.F., A. Larsson); and Dipartimento di Elettronica, Informazione e Bioingegneria, Politecnico di Milano, Milan, Italy (A. LoMauro, S.G., R.P., A.A.).

Abstract

Abstract Background: Both superimposed high-frequency jet ventilation (SHFJV) and single-frequency (high-frequency) jet ventilation (HFJV) have been used with success for airway surgery, but SHFJV has been found to provide higher lung volumes and better gas exchange than HFJV in unobstructed airways. The authors systematically compared the ventilation efficacy of SHFJV and HFJV at different ventilation frequencies in a model of tracheal obstruction and describe the frequency and obstruction dependence of SHFJV efficacy. Methods: Ten anesthetized animals (weight 25 to 31.5 kg) were alternately ventilated with SHFJV and HFJV at a set of different fHF from 50 to 600 min−1. Obstruction was created by insertion of interchangeable stents with ID 2 to 8 mm into the trachea. Chest wall volume was measured using optoelectronic plethysmography, airway pressures were recorded, and blood gases were analyzed repeatedly. Results: SHFJV provided greater than 1.6 times higher end-expiratory chest wall volume than HFJV, and tidal volume (VT) was always greater than 200 ml with SHFJV. Increase of fHF from 50 to 600 min−1 during HFJV resulted in a more than 30-fold VT decrease from 112 ml (97 to 130 ml) to negligible values and resulted in severe hypoxia and hypercapnia. During SHFJV, stent ID reduction from 8 to 2 mm increased end-expiratory chest wall volume by up to 3 times from approximately 100 to 300 ml and decreased VT by up to 4.2 times from approximately 470 to 110 ml. Oxygenation and ventilation were acceptable for 4 mm ID or more, but hypercapnia occurred with the 2 mm stent. Conclusion: In this in vivo porcine model of variable severe tracheal stenosis, SHFJV effectively increased lung volumes and maintained gas exchange and may be advantageous in severe airway obstruction.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Anesthesiology and Pain Medicine

Reference36 articles.

1. Outcomes of tracheobronchial stents in patients with malignant airway disease.;Ann Thorac Surg,2005

2. Exercise capacity, lung function, and quality of life after interventional bronchoscopy.;J Thorac Oncol,2011

3. Anaesthesia and tracheobronchial stenting for central airway obstruction in adults.;Br J Anaesth,2003

4. Superimposed high-frequency jet ventilation (SHFJV) for endoscopic laryngotracheal surgery in more than 1500 patients.;Br J Anaesth,2006

5. [Anesthesia in acute respiratory tract obstructions caused by high degree laryngeal and tracheobronchial stenoses].;Anasthesiol Intensivmed Notfallmed Schmerzther,1994

Cited by 17 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献