Validation of the Hingmed DBP-01P clinical automatic blood pressure monitor provided with wide-range cuff, evaluated in a general population according to the ISO 81060-2:2018 protocol

Author:

Fania Claudio1,Giletto Antonella1,Zanello Valentina1,Palatini Paolo2

Affiliation:

1. Villa Maria Hospital, Department of Medicine, Via Delle Melette, Padova

2. Studium Patavinum, Department of Medicine, University of Padova, Padua, Italy

Abstract

To determine the accuracy of the Hingmed DBP-01P monitor for clinical automatic blood pressure (BP) measurement according to the International Organization for Standardization (ISO) 81060-2 2018 protocol. The DBP-01P was tested in 85 subjects from the general population (mean age, 69.9 years; 43 men) using a wide-range cuff for arm circumferences from 17 to 42 cm. The mean device–observer difference was 1.32 ± 3.18 mmHg for SBP and 0.78 ± 3.15 mmHg for DBP. These data were in agreement with criterion 1 of the protocol standard requirements (≤5 ± 8 mmHg). Also, criterion 2 was satisfied with the SDs of the 85 participants being well below the maximum values required by the protocol (6.80 and 6.89 mmHg for SBP and DBP pressure, respectively). These data show that the Hingmed BP monitor DBP-01P satisfied the ISO 81060-2:2018 standard requirements for a general population across a wide range of arm sizes using a single wide-range cuff.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Advanced and Specialized Nursing,Assessment and Diagnosis,Cardiology and Cardiovascular Medicine,General Medicine,Internal Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3