Validation of the Meditech ABPM-06 24-hour blood pressure monitoring system in a pediatric population according to International Organization for Standardization Protocol 81060-2:2018

Author:

Jakab Andrea E.1,Horváth Edit1,Molnár Dorottya1,Bukva Mátyás1,Bereczki Csaba1

Affiliation:

1. Department of Pediatrics and Pediatric Health Center, Albert Szent-Györgyi Health Centre, University of Szeged Albert Szent-Györgyi Medical School, Korányi fasor, Szeged, Hungary

Abstract

Objective Ambulatory blood pressure monitoring (ABPM) devices play a crucial role in diagnosing hypertension, not only in adults but also in pediatric patients. ABPM-06, the latest oscillometric device from Meditech Ltd. (Budapest, Hungary), is the focus of this study. The objective was to validate the ABPM-06 device using the International Organization for Standardization (ISO) 81060-2 : 2018 standard. Methods A total of 86 healthy patients, consisting of 34 males and 52 females, aged between 3 and 17 years, were enrolled in this study. During the ambulatory phase, a total of 50 patients were enrolled, with 35 patients falling within the 3- to 12-year-old age range and 15 patients aged between 12 and 17 years. Additionally, for the dynamic test, 36 patients were selected, comprising of 10 individuals aged 3–12 years and 26 patients aged 12–17 years. These patients were recruited from the outpatient clinics of the Department of Pediatrics at Albert Szent-Györgyi University in Szeged, Hungary. The validation process involved utilizing the same-arm sequence protocol, both in resting positions and during stress testing. Results The ABPM-06 performed well in both clinical and ambulatory validations. In terms of validation criterion 1, the mean ± SD of the differences between the test device and reference blood pressure readings was −1.3 ± 3.5  mmHg for systolic and −0.1  ± 2.3 mmHg for diastolic, in children under the age of 12 years. For those over the age of 12 years, the mean ± SD of the differences was −2.8 ± 4.6 mmHg for systolic and −0.5 ± 2.7 mmHg for diastolic. Regarding the ambulatory validation, for children under 12 years old, the mean ± SD of the differences was −1.3  ± 3.5  mmHg for systolic and −0.1  ± 2.3  mmHg for diastolic. In the age group above 12 years, the mean ± SD of the differences was −2.8 ± 4.6 mmHg for systolic and −0.5 ± 2.7 mmHg for diastolic. Both tests successfully met the established criteria regarding the mean and SD values of the differences between the device readings and the observed SBP and DBP measurements. Conclusion The ABPM-06 oscillometric device fully adheres to the ISO 81060-2 : 2018 standard requirements for ABPM determination in the pediatric population (ages 3–17 years). Consequently, this ABPM device proves to be suitable for effectively managing hypertension in children and adolescents.

Publisher

Ovid Technologies (Wolters Kluwer Health)

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3