Abstract
Abstract
PURPOSE
Optical Calorimetry (OC) is based on interferometry and provides a direct measurement of spatially resolved absorbed dose to water by measuring refractive index changes induced by radiation. The purpose of this work was to optimize and characterize in software an OC system tailored for ultra-high dose rate applications and to build and test a prototype in a clinical environment.
METHODS
A radiation dosimeter using the principles of OC was designed in optical modelling software. Traditional image quality instruments, fencepost and contrast phantoms, were utilized in both software and experimentally in a lab environment to investigate noise reduction techniques and to test the spatial and dose resolution of the system. Absolute dose uncertainty was assessed by measurements in a clinical 6 MV flattening filter free (FFF) photon beam with dose rates in the range 0.2–6 Gy/s achieved via changing the distance from the source.
RESULTS
Design improvements included: equalizing the pathlengths of the interferometer, isolating the system from external vibrations and controlling the system’s internal temperature as well as application of mathematical noise reduction techniques. Simulations showed that these improvements should increase the spatial resolution from 22 to 35 lp/mm and achieve a minimum detectable dose of 0.2 Gy, which was confirmed experimentally. In the FFF beam, the absolute dose uncertainty was dose rate dependent and decreased from 2.5 ± 0.8 Gy to 2.5 ± 0.2 Gy for dose rates of 0.2 and 6 Gy/s, respectively.
CONCLUSIONS
A radiation dosimeter utilizing the principles of OC was developed and constructed. Optical modelling software and image quality phantoms allowed for iterative testing and refinement. The refined OC system proved capable of measuring absorbed dose to water in a linac generated photon beam. Reduced uncertainty at higher dose rates indicates the potential for OC as a dosimetry system for high dose rate techniques such as microbeam and ultra-high dose-rate radiotherapy.