COMPARTMENT MONITORING IN TIBIAL FRACTURES

Author:

McQueen M. M.1,Court-Brown C. M.1

Affiliation:

1. Orthopaedic Trauma Unit, Royal Infirmary of Edinburgh NHS Trust, Lauriston Place, Edinburgh EH3 9YW, UK.

Abstract

We made a prospective study of 116 patients with tibial diaphyseal fractures who had continuous monitoring of anterior compartment pressure for 24 hours. Three patients had acute compartment syndrome (2.6%). In the first 12 hours of monitoring, 53 patients had absolute pressures over 30 mmHg and 30 had pressures over 40 mmHg, with four higher than 50 mmHg. Only one patient had a differential pressure (diastolic minus compartment pressure) of less than 30 mmHg; he had a fasciotomy. In the second 12-hour period 28 patients had absolute pressures over 30 mmHg and seven over 40 mmHg. Only two had differential pressures of less than 30 mmHg; they had fasciotomies. None of our 116 patients had any sequelae of the compartment syndrome at their latest review at least six months after injury. A threshold for decompression of 30 mmHg would have indicated that 50 patients (43%) would have required fasciotomy, and at a 40 mmHg threshold 27 (23%) would have been considered for an unnecessary fasciotomy. In our series, the use of a differential pressure of 30 mmHg as a threshold for fasciotomy led to no missed cases of acute compartment syndrome. We recommended that decompression should be performed if the differential pressure level drops to under 30 mmHg.

Publisher

British Editorial Society of Bone & Joint Surgery

Subject

Orthopedics and Sports Medicine,Surgery

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3