CORROSION OF ORTHOPAEDIC IMPLANTS

Author:

Scales John T.,Winter G. D.,Shirley H. T.

Abstract

1. A total of 564 metal components from 109 patients have been examined. 2. Corrosion was detected on 228 components. 3. Most implants were removed for reasons other than corrosion. 4. In at least twelve cases corrosion was the reason for removal of the implant. 5. No corrosion of cobalt-chromium-molybdenum implants has been detected by the methods described in this paper. 6. Corrosion of ferrous alloy implants occurs in the human body. 7. The most common site for corrosion is the junction between components of implants. 8. The most corrosion-resistant type of ferrous alloy for implant uses is 18/8/Mo steel, which falls within specification En58.J of the British Standards Institute, and specification 316/317 of the American Iron and Steel Institute. 9. About 13 per cent of components removed (and by no means only when something was amiss) showed "face" corrosion when all the components of the implant were made of 18/8/Mo steel. 10. Four per cent of 18/8/Mo components of implants showed corrosion at sites other than the interface between components. 11. On the basis of corrosion resistance it is preferable to use cobalt-chromium-molybdenum alloys for implants that are to be left in the body for an indefinite period. 12. The corrosion resistance of the high alloy steels examined (18/8, 18/8/Ti, 18/8/Mo) does not appear to be related to hardness. 13. The marking of components, either by punching or by electrolytic methods, has not predisposed to corrosion. 14. All hollow 18/8/Mo implants should have a clean and metallurgically satisfactory internal surface. 15. The figures in this report do not permit a full statistical survey of corrosion in surgical implants because it has not been possible to examine a satisfactorily random sample. Many patients cannot be followed up and others die with the implant still in the body.

Publisher

British Editorial Society of Bone & Joint Surgery

Subject

Orthopedics and Sports Medicine,Surgery

Cited by 93 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3